Abstract

Despite the success of total knee arthroplasty (TKA), malalignment continues to be a problem which often leads to post-operative complications. The aim of this study was to investigate the accuracy of a novel, imageless, optical surgical navigation tool to assist with the alignment of femoral and tibial cuts performed during total knee arthroplasty. Six board-certified orthopedic surgeons performed TKA procedures on 9 cadavers (17 knees total), using a novel, imageless navigation system (Intellijoint KNEE, Intellijoint Surgical). Varus/valgus, femoral flexion, tibial slope, and rotation measurements from the device were compared with angular measurements calculated from post-operative computed tomography (CT) images. Navigation measurements were highly correlated with those obtained from CT scan in all three axes. For the femoral cuts, the absolute mean difference in varus/valgus was 0.83° (SD 0.46°, r = 0.76), in flexion was 1.91° (SD 1.16°, r = 0.85), and in rotation was 1.29° (SD 1.01°, r = 0.88) relative to Whiteside’s line and 0.97° (SD 0.56°, r = 0.81) relative to the posterior condylar axis. For the tibia, the absolute mean difference in varus/valgus was 1.08° (SD 0.64°, r = 0.85), anterior/posterior slope was 2.78° (SD 1.40°, r = 0.60), and rotation was 2.98° (SD 2.54°, r = 0.79). Intraoperative monitoring with the imageless navigation tool accurately measures femoral and tibial cuts in TKA and may help to increase component alignment.

Keyphrases: Accuracy, alignment, Computer Assisted Navigation, Total knee arthroplasty

In: Ferdinando Rodriguez Y Baena, Joshua W Giles and Eric Stindel (editors). Proceedings of The 20th Annual Meeting of the International Society for Computer Assisted Orthopaedic Surgery, vol 5, pages 158--161

Links:
BibTeX entry
@inproceedings{CAOS2022:cadaveric_investigation_of_accuracy,
  author    = {Ran Schwarzkopf and Brian Culp and Michael Bradley and Emily McIntosh},
  title     = {A cadaveric investigation of the accuracy of a new, computer-assisted navigation system for total knee arthroplasty: A comparison with computed tomography imaging},
  booktitle = {Proceedings of The 20th Annual Meeting of the International Society for Computer Assisted Orthopaedic Surgery},
  editor    = {Ferdinando Rodriguez Y Baena and Joshua W Giles and Eric Stindel},
  series    = {EPiC Series in Health Sciences},
  volume    = {5},
  pages     = {158--161},
  year      = {2022},
  publisher = {EasyChair},
  bibsource = {EasyChair, https://easychair.org},
  issn      = {2398-5305},
  url       = {https://easychair.org/publications/paper/lX74F},
  doi       = {10.29007/qmj5}}
Download PDFOpen PDF in browser